Copyright 2024 - Custom text here

If you are interested in using the Tecan Safire 2 Multi-Mode Plate Reader or if you wish to consult us for your research, please contact:

Dr. Radim Osička

E-mail: This email address is being protected from spambots. You need JavaScript enabled to view it.

Phone: +420 241 062 770

 

Tecan Safire 2 Plate Reader Specifications:

Plate formats 6 to 1536 well plates
Temperature control Ambient + 4 °C to 42 °C
Shaking linear, orbital, 3 velocities
Fluorescence Intensity  
Wavelength selection Monochromator
Wavelength range Excitation: 230 – 850 nm, emission: 280 – 850 nm
Bandwidth Adjustable from 2.5 nm – 10 nm below 280 nm, 5 nm – 20 nm else
Wavelength accuracy ± 1 nm
Wavelength precision ± 1 nm
Sensitivity – top reading  
– 384 well plate 0.2 fmol / well Fluorescein
Sensitivity – bottom reading  
– 384 well plate 0.6 fmol / well Fluorescein
Time Resolved Fluorescence  
Sensitivity* – 384 well plate 60 amol / well Europium
Precision – 384 well plate 3 mP standard deviation @ 1nM Fluorescein
Absorbance  
Wavelength range 230 – 1000 nm
Bandwidth 2.5 nm below 280nm, 5 nm else
Wavelength accuracy ± 1 nm
Wavelength precision ± 1 nm
Measurement range 0 – 4OD
Accuracy @ 492 nm 0 – 2OD: ± 1% and 0.006OD
Precision @ 492 nm 0 – 2OD: ± 0.5% and 0.005OD
Fluorescence Absorbance FP  
384 well plate 25 sec 25 sec 45 sec
1536 well plate 37 sec 27 sec 108 sec
Interface connections RS 232 serial & parallel port
Software Driven YES
Power 115-120V / 220-240V 50/60Hz
Dimensions 22" x 21" x 11"
Weight 88 lb