Tecan Safire 2 Plate Reader Specifications:
Plate formats | 6 to 1536 well plates |
Temperature control | Ambient + 4 °C to 42 °C |
Shaking | linear, orbital, 3 velocities |
Fluorescence Intensity | |
Wavelength selection | Monochromator |
Wavelength range | Excitation: 230 – 850 nm, emission: 280 – 850 nm |
Bandwidth | Adjustable from 2.5 nm – 10 nm below 280 nm, 5 nm – 20 nm else |
Wavelength accuracy | ± 1 nm |
Wavelength precision | ± 1 nm |
Sensitivity – top reading | |
– 384 well plate | 0.2 fmol / well Fluorescein |
Sensitivity – bottom reading | |
– 384 well plate | 0.6 fmol / well Fluorescein |
Time Resolved Fluorescence | |
Sensitivity* – 384 well plate | 60 amol / well Europium |
Precision – 384 well plate | 3 mP standard deviation @ 1nM Fluorescein |
Absorbance | |
Wavelength range | 230 – 1000 nm |
Bandwidth | 2.5 nm below 280nm, 5 nm else |
Wavelength accuracy | ± 1 nm |
Wavelength precision | ± 1 nm |
Measurement range | 0 – 4OD |
Accuracy @ 492 nm | 0 – 2OD: ± 1% and 0.006OD |
Precision @ 492 nm | 0 – 2OD: ± 0.5% and 0.005OD |
Fluorescence Absorbance FP | |
384 well plate | 25 sec 25 sec 45 sec |
1536 well plate | 37 sec 27 sec 108 sec |
Interface connections | RS 232 serial & parallel port |
Software Driven | YES |
Power | 115-120V / 220-240V 50/60Hz |
Dimensions | 22" x 21" x 11" |
Weight | 88 lb |